Description: Alaris Infusion Pump 8100 Bezel (plastic piece only), Part: TIPA-8100-4410

Tenacore LLC Recall 87470 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1332-2021
Event ID87470
Event DescriptionAlaris Infusion Pump 8100 Bezel (plastic piece only), Part: TIPA-8100-4410
Product TypeDevices
DistributionUS distribution to states of: WA, AZ, NY, TX, LA, CA, NV, ME, MN, DE, NC, FL, MO, KY, TN, and PA
Quantity1494
Recall ReasonBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Device Classification20210414
Device Code InfoParts distributed between July 10, 2020 and January 22, 2021. Time stamp at bezel rear side shows 5 or 6 SO Numbers: 271111, 263873, 264442, 264890, 268606, 269930, 270760, 270936, 271211, 272860, 267603, 268936, 269258, 269291, 272496, 272225, 271408, 267392
Center Classification Date20210402
Recall Initiation Date20210225
Recalling FirmTenacore LLC