Product Safety Recalls

Product Recall Tracker

Coloplast Manufacturing US, LLC Recall 91800

Description: Titan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis

Coloplast Manufacturing US, LLC Recall 91800 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1360-2023
Event ID91800
Event DescriptionTitan Pump Assembly, Catalog Number 5177701400; inflatable penile Prosthesis
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of Turkey, Italy, Germany, Belgium, France, United Kingdom, Spain, Portugal.
Quantity9 units
Recall ReasonA decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Device Classification20230419
Device Code InfoUDI/DI 05708932319621, Lot Numbers: 8904177, 8849617
Center Classification Date20230412
Recall Initiation Date20230301
Recalling FirmColoplast Manufacturing US, LLC
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.