Coloplast Manufacturing US, LLC Recall 91800

Description: Titan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis

Coloplast Manufacturing US, LLC Recall 91800 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1366-2023
Event ID91800
Event DescriptionTitan NB Scrotal Zero Ang 18cm, Catalog Number EN88182400; inflatable penile Prosthesis
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of Turkey, Italy, Germany, Belgium, France, United Kingdom, Spain, Portugal.
Recall ReasonA decrease in wall thickness has the potential for a premature pump failure, compared to a pump with standard wall thickness, based on the number of interactions with the pump to inflate and deflate the penile prosthesis.
Device Classification20230419
Device Code InfoUDI/DI 05708932072649, Lot Numbers: 8887622, 8965351
Center Classification Date20230412
Recall Initiation Date20230301
Recalling FirmColoplast Manufacturing US, LLC
Initial Notification Letter
Similar To