Covidien, LP Recall 87475
Description: Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT
Covidien, LP Recall 87475 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1383-2021 |
Event ID | 87475 |
Event Description | Covidien Signia Small Diameter Curved Tip Intelligent Reload 45 mm Vascular/Thin 8 mm - Long Item Code: SIGSDL45CTVT |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom. |
Quantity | 5928 units WW: 4212 US 1716 OUS |
Recall Reason | Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure. |
Device Classification | 20210421 |
Device Code Info | Lot Numbers: N0A1038UY N0E0016UY N0F0616UY N0K0934UY N0B0493UY N0F0254UY N0F0817UY N0M0334UY N0C0436UY N0F0518UY N0G0388UY N9K1095UY N0D0731UY N0F0615UY N0J0859UY N9L0797UY UDI: 10884521741843 20884521741840 |
Center Classification Date | 20210409 |
Recall Initiation Date | 20210304 |
Recalling Firm | Covidien, LP |
Initial Notification | Letter |
Similar To |