Covidien Recall 87542
Description: Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094
Covidien Recall 87542 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1411-2021 |
Event ID | 87542 |
Event Description | Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094 |
Product Type | Devices |
Distribution | Domestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador, Guatemala, Jamaica, Mexico, Panama, Peru, Puerto Rico, Trinidad And Tobago, Austria, Belgium, Bosnia And Herzegovina, Canary Islands, Cyprus, Denmark, Equatorial Guinea, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Kosovo, Kuwait, Lebanon, Mauritius, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, South Africa, Spain, Swaziland, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom. |
Quantity | 59,828 devices |
Recall Reason | Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector. |
Device Classification | 20210421 |
Device Code Info | UDI-Device Identifier (GTIN/UPN): 10884523000887, 20884523000884 UDI-Device Identifier (GTIN/UPN): 10884523000900, 20884523000907 UDI-Device Identifier (GTIN/UPN): 10884523000917, 20884523000914 |
Center Classification Date | 20210412 |
Recall Initiation Date | 20210305 |
Recalling Firm | Covidien |
Initial Notification | Letter |
Similar To |