Life Technologies Corporation Recall 94122
Description: Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The Ion PGM Dx Instrument System is not intended for whole genome or de novo sequencing Product Name Installer /SKU /Updater SKU: (1) Torrent Suite Dx Software 5.14 -A52422/ A52417 (2) Torrent Suite Dx Software 5.12.5- A46166/ A46167 (3)Torrent Suite Dx Software 5.8- A36601/ A36602 (4)Torrent Suite Dx Software 5.6.4- A33178/ A33178 (5)Torrent Suite Dx Software 5.0- A29166/ A29166
Life Technologies Corporation Recall 94122 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1426-2024 |
Event ID | 94122 |
Event Description | Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems-IVD intended for targeted sequencing of human genomic DNA (gDNA) from peripheral whole-blood samples and DNA and RNA extracted from formalin-fixed, paraffin-embedded (FFPE) samples. The Ion PGM Dx Instrument System is not intended for whole genome or de novo sequencing Product Name Installer /SKU /Updater SKU: (1) Torrent Suite Dx Software 5.14 -A52422/ A52417 (2) Torrent Suite Dx Software 5.12.5- A46166/ A46167 (3)Torrent Suite Dx Software 5.8- A36601/ A36602 (4)Torrent Suite Dx Software 5.6.4- A33178/ A33178 (5)Torrent Suite Dx Software 5.0- A29166/ A29166 |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of CA, IN, NC OR, TN. |
Quantity | 28 units |
Recall Reason | Torrent Suite Dx Software versions 5.14 and earlier used in connection with Ion PGM DX Systems exploitation of the vulnerability by a threat actor may allow them to alter settings, configurations, software, or data on the instrument |
Device Classification | 20240410 |
Device Code Info | UDI: 10190302017848;10190302016810 Torrent Suite Dx Software versions 5.14 and earlier |
Center Classification Date | 20240329 |
Recall Initiation Date | 20240207 |
Recalling Firm | Life Technologies Corporation |
Initial Notification | Letter |
Similar To |