Aesculap Implant Systems LLC Recall 90502
Description: AESCULAP S4 Cervical Cross Connector L24mm, S4C CROSS CONNECTOR FIXED 24MM, product code SWT113T. Used in spinal fusion surgery.
Aesculap Implant Systems LLC Recall 90502 Information
Status | Terminated |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1444-2022 |
Event ID | 90502 |
Event Description | AESCULAP S4 Cervical Cross Connector L24mm, S4C CROSS CONNECTOR FIXED 24MM, product code SWT113T. Used in spinal fusion surgery. |
Product Type | Devices |
Distribution | United States Nationwide distribution in the states of CA, CO, IL, IN, MI, NC, NV, OH, PA, SC, TX & WY. |
Quantity | 24 |
Recall Reason | Insufficient clamping force of the cross connectors in the spinal surgical system which may lead to movement of the connectors on the rods and/or microparticle abrasion. |
Device Classification | 20220727 |
Device Code Info | UDI-DI: 04038653362301 |
Center Classification Date | 20220721 |
Recall Initiation Date | 20200908 |
Recalling Firm | Aesculap Implant Systems LLC |
Initial Notification | Letter |
Similar To |