Hamilton Medical AG Recall 90576

Description: HAMILTON-C6 Intensive Care Ventilator, REF: 160021

Hamilton Medical AG Recall 90576 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1525-2022
Event ID90576
Event DescriptionHAMILTON-C6 Intensive Care Ventilator, REF: 160021
Product TypeDevices
DistributionUS nationwide distribution including Puerto Rico.
Quantity358
Recall ReasonVentilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed
Device Classification20220831
Device Code InfoUDI-DI: 07630002808590, Serial Numbers: 5663, 6995, 6907, 6952, 7177, 7772, 6932, 6157, 8658, 7307, 5939, 6858, 6231, 8620, 6349, 8356, 6480, 8322, 8306, 9100, 9101, 9099, 9098, 6219, 7894, 7861, 8003, 6723, 7802, 4833, 7265, 7980, 7666, 7382, 6774, 7358, 7719, 6896, 6925, 9073, 9062, 9298, 9294, 9301, 9292, 9293, 9290, 9300, 9299, 9291, 9231, 9132, 9108, 8310, 8604, 8255, 8599, 7156, 8313, 8344, 8590, 8087, 8348, 8490, 8151, 7174, 8594, 8477, 8122, 8063, 9353, 9230, 9329, 9243, 9235, 9236, 9337, 9228, 9224, 9229, 9225, 9103, 8413, 6149, 8178, 8401, 7827, 8406, 8409, 7826, 6416, 7831, 6280, 9128, 9260, 9257, 9125, 9120, 9259, 9258, 9354, 8325, 8297, 7777, 6212, 6516, 6871, 8099, 6562, 8258, 5679, 5681, 7808, 7115, 8601, 7349, 7130, 6658, 7581, 7819, 8346, 6537, 8235, 7410, 6868, 8049, 8281, 7966, 7787, 9241, 9242, 9054, 9075, 9131, 9121, 9137, 9134, 9133, 9118, 6350, 7122, 7117, 6790, 7614, 8121, 7884, 8102, 8720, 8715, 8726, 8718, 8719, 9248, 9244, 9127, 9138, 9126, 8256, 8371, 9107, 9109, 9036, 9110, 9069, 9063, 9067, 8097, 9129, 9130, 9459, 8723, 8722, 8713, 8702, 8725, 8714, 8727, 8706, 8517, 8717, 7822, 8148, 7695, 8261, 7165, 6944, 7420, 6937, 7150, 6953, 7201, 6965, 7004, 7196, 8849, 8597, 8850, 9106, 9192, 9239, 9237, 9233, 8311, 7898, 7893, 6641, 6505, 6075, 7986, 8005, 7888, 7887, 9535, 9579, 9508, 9576, 9553, 9577, 9578, 9545, 9582, 9590, 9575, 7845, 7740, 8082, 8070, 6671, 7608, 6566, 8535, 8616, 7821, 6579, 7586, 7994, 7066, 7138, 9102, 9105, 14784, 8603, 8846, 8844, 8847, 8845, 8115, 7113, 7809, 7938, 7993, 7252, 8568, 7754, 7033, 8052, 6829, 8249, 9077, 9080, 9076, 9051, 9064, 9081, 9050, 9071, 9091, 7836, 7656, 9124, 9462, 9297, 9325, 9303, 8866, 9348, 9352, 9324, 9227, 6093, 6159, 6274, 6276, 6309, 6320, 6327, 6464, 6512, 6545, 6674, 6789, 7136, 7173, 7191, 7273, 7279, 7422, 7432, 7569, 7572, 7587, 7644, 7676, 7782, 7783, 7784, 7786, 7788, 7789, 7790, 7792, 7793, 7794, 7795, 7796, 7797, 7798, 7799, 7801, 7806, 7807, 7812, 7813, 7837, 7901, 7923, 7969, 8182, 8343, 8394, 8395, 8399, 8420, 8421, 8422, 8423, 8424, 8497, 8519, 8593, 8661, 8668, 8670, 8673, 8675, 8676, 8678, 8707, 8709, 8712, 9028, 9029, 9032, 9033, 9042, 9043, 9044, 9046
Center Classification Date20220819
Recall Initiation Date20220627
Recalling FirmHamilton Medical AG
Initial Notification Letter