Greiner Bio-One North America, Inc. Recall 87660
Description: VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% 13x75 blue cap-white ring, sandwich tube, non-ridged - Product Usage: indicated for use in collection and transport of venous whole blood for coagulation testing.
Greiner Bio-One North America, Inc. Recall 87660 Information
Status | Completed |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1534-2021 |
Event ID | 87660 |
Event Description | VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% 13x75 blue cap-white ring, sandwich tube, non-ridged - Product Usage: indicated for use in collection and transport of venous whole blood for coagulation testing. |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of KY, TX, LA, OH, AR, IL, SC, PA, FL, NC, AZ, CA, MI, NE, KS, MI, MO, NY, and NJ. |
Quantity | 1,049,700 units |
Recall Reason | VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% may show insufficient draw volume of the tubes, which could produce erroneous results for prothrombin time, activated partial thromboplastin time, and fibrinogen assays. |
Device Classification | 20210512 |
Device Code Info | Item Number: 454322; Lot # A21023B5; UDI: (Case) 39120017574485 and (Rack) 29120017574488; Expiration Date: 02/06/2022 Item Number: 454332; Lot # A210338Q; UDI: (Case) 39120017574607 and (Rack) 29120017574600; Expiration Date: 03/01/2022 Item Number: 454334; Lot # A210233V; UDI: (Case) 39120017574539 and (Rack) 29120017574532; Expiration Date: 02/01/2022 Item Number: 454334; Lot # A21023XW; UDI: (Case) 39120017574539 and (Rack) 29120017574532; Expiration Date: 02/13/2022 |
Center Classification Date | 20210506 |
Recall Initiation Date | 20210329 |
Recalling Firm | Greiner Bio-One North America, Inc. |
Initial Notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |