Infusion Pump Repair Recall 87657

Description: Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part

Infusion Pump Repair Recall 87657 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1544-2021
Event ID87657
Event DescriptionAlaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part
Product TypeDevices
DistributionCalifornia
Quantity15
Recall ReasonBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Device Classification20210519
Device Code InfoDevices distributed between 01/15/2021 and 03/14/2021. Sale Order #/Serial #: 1599930872/ 12335834, 13617863, 13447266, 13089105, 12333728, 13616451, 13618575, 12729374; 82943/ 4122252, 4122320; 01112101A/ 12638236, 13813590, 12639090, 13744738, 9951014
Center Classification Date20210510
Recall Initiation Date20210326
Recalling FirmInfusion Pump Repair