Ortho-Clinical Diagnostics, Inc. Recall 92055
Description: VITROS 5600 Integrated System- For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents Product Code: 6802413
Ortho-Clinical Diagnostics, Inc. Recall 92055 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1556-2023 |
Event ID | 92055 |
Event Description | VITROS 5600 Integrated System- For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents Product Code: 6802413 |
Product Type | Devices |
Distribution | Nationwide including Puerto Rico and Bermuda Foreign: Australia Belgium Brazil Canada Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore Spain Sweden The Netherlands United Kingdom |
Quantity | 4895 units |
Recall Reason | MicroTip Pack Opener Assembly Potentially Not Removing or Replacing MicroTip Pack Caps on VITROS¿ 5600 and XT 7600 Integrated Systems, results may be delayed results if a pack becomes unusable |
Device Classification | 20230510 |
Device Code Info | UDI: 10758750002740 Serial Number Range: J56000143- J56004477 |
Center Classification Date | 20230504 |
Recall Initiation Date | 20230310 |
Recalling Firm | Ortho-Clinical Diagnostics, Inc. |
Initial Notification | Letter |