ACTIM OY Recall 93845

Description: Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane

ACTIM OY Recall 93845 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1592-2024
Event ID93845
Event DescriptionActim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane
Product TypeDevices
DistributionUS Nationwide distribution in the state of Connecticut.
Quantity6471 kits
Recall ReasonInterference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result.
Device Classification20240424
Device Code InfoUDI/DI 06430030990608, All serial numbers in distribution
Center Classification Date20240417
Recall Initiation Date20231123
Recalling FirmACTIM OY