ACTIM OY Recall 93845
Description: Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane
ACTIM OY Recall 93845 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1592-2024 |
Event ID | 93845 |
Event Description | Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane |
Product Type | Devices |
Distribution | US Nationwide distribution in the state of Connecticut. |
Quantity | 6471 kits |
Recall Reason | Interference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result. |
Device Classification | 20240424 |
Device Code Info | UDI/DI 06430030990608, All serial numbers in distribution |
Center Classification Date | 20240417 |
Recall Initiation Date | 20231123 |
Recalling Firm | ACTIM OY |