Product Safety Recalls

Product Recall Tracker

Roche Diagnostics Operations, Inc. Recall 87705

Description: Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

Roche Diagnostics Operations, Inc. Recall 87705 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1595-2021
Event ID87705
Event DescriptionElecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
Product TypeDevices
DistributionUS Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.
Quantity10, 489 kits
Recall ReasonRecall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.
Device Classification20210526
Device Code InfoUDI: 04015630939381
Center Classification Date20210514
Recall Initiation Date20210407
Recalling FirmRoche Diagnostics Operations, Inc.
Initial Notification FAX
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.