RAYSEARCH LABORATORIES AB Recall 87868

Description: RayStation standalone software radiation therapy treatment planning system. Model Number: 8A, 8B, 9A, 9B, 10A, 10B including some service packs

RAYSEARCH LABORATORIES AB Recall 87868 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1598-2021
Event ID87868
Event DescriptionRayStation standalone software radiation therapy treatment planning system. Model Number: 8A, 8B, 9A, 9B, 10A, 10B including some service packs
Product TypeDevices
DistributionDomestic: TN
Recall ReasonWhen using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose tracking and the fractions that are included in the dose summation. The issue only affects RayTreat users and only the Total dose option. Accumulation and other options in Dose tracking are correct.
Device Classification20210526
Device Code InfoSoftware version number: 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 , 10.1.0.613
Center Classification Date20210514
Recall Initiation Date20210407
Recalling FirmRAYSEARCH LABORATORIES AB