Beckman Coulter, Inc. Recall 92056

Description: DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis

Beckman Coulter, Inc. Recall 92056 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1598-2023
Event ID92056
Event DescriptionDxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis
Product TypeDevices
DistributionWorldwide Distribution
Quantity355 units
Recall ReasonBeckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.
Device Classification20230524
Device Code InfoUDI/DI 15099590748654, All analyzers with Windows 10, Software Versions 8.5, 8.5.1 and 8.6
Center Classification Date20230517
Recall Initiation Date20230405
Recalling FirmBeckman Coulter, Inc.
Initial Notification Letter