ROUTE 92 MEDICAL INC Recall 94270

Description: Route 92 Medical, Tenzing 7 Delivery Catheter, REF 1005, Rx Only, Sterile EO. Used for introduction of interventional devices into the neurovasculature.

ROUTE 92 MEDICAL INC Recall 94270 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1613-2024
Event ID94270
Event DescriptionRoute 92 Medical, Tenzing 7 Delivery Catheter, REF 1005, Rx Only, Sterile EO. Used for introduction of interventional devices into the neurovasculature.
Product TypeDevices
DistributionUS: AL AZ CA DE FL GA IA IL IN KY LA MA MD MI MN MO NC ND NJ NY OH OR PA RI SD TN TX UT VA WV OUS: New Zealand
Quantity200 units
Recall ReasonTheir is the potential for distal tip separation at the proximal marker band of the delivery catheters
Device Classification20240501
Device Code InfoUDI: 00853799007534/ Lot # 23033001, 23073101, 23070305, 23081707
Center Classification Date20240424
Recall Initiation Date20240308
Recalling FirmROUTE 92 MEDICAL INC