PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Recall 92165
Description: Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Azurion 3M12: 722063, 722221; (2) Azurion 3M15: 722064, 722222; (3) Azurion SM12: 722227; (4) Azurion SM20: 722228; (5) Azurion 7B12/12: 722067, 722225; (6) Azurion 7820/15: 722068, 722226; (7) Azurion 7M12: 722078, 722223; (8) Azurion 7M20 : 722079, 722224
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Recall 92165 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1647-2023 |
Event ID | 92165 |
Event Description | Philips Azurion System with software versions: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6. Models/Model Number: (1) Azurion 3M12: 722063, 722221; (2) Azurion 3M15: 722064, 722222; (3) Azurion SM12: 722227; (4) Azurion SM20: 722228; (5) Azurion 7B12/12: 722067, 722225; (6) Azurion 7820/15: 722068, 722226; (7) Azurion 7M12: 722078, 722223; (8) Azurion 7M20 : 722079, 722224 |
Product Type | Devices |
Distribution | Nationwide Foreign: Algeria Argentina Armenia Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Brazil Bulgaria Cambodia Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cyprus Czech Republic Denmark Dominican Rep Ecuador Egypt Estonia Finland France French Polynesia Germany Greece Hong Kong Hungary India Indonesia Iran Iraq Ireland Israel Italy Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Latvia Lebanon Lithuania Macao Macedonia Malaysia Mexico Morocco Myanmar Nepal Netherlands New Zealand Nicaragua Nigeria Norway Oman Pakistan Palestine Palestine, State of Panama Peru Philippines Poland Portugal Puerto Rico Qatar Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa South Korea Spain Sri Lanka Sweden Switzerland Taiwan Thailand Tunisia Turkey Turkmenistan Ukraine United Kingdom Utd.Arab.Emir. Uzbekistan Vietnam Yemen |
Quantity | USA: 1063 R.O.W. 2136 |
Recall Reason | Potential Loss of X-ray-When the full disk capacity is reached, X-Ray functionality will cease to be available there will be a sudden interruption of the procedure |
Device Classification | 20230607 |
Device Code Info | UDI-DI: System product name Model number UDI (1)Azurion 3M12: 722063 (01)00884838085275(21) 722221 (01)00884838099203(21); (2) Azurion 3M15: 722064 (01)00884838085282(21) 722222 (01)00884838099210(21); (3) Azurion SM12: 722227 (01)00884838099227(21); (4) Azurion SM20: 722228 (01)00884838099234(21); (5) Azurion 7B12/12: 722067 (01)00884838085350(21) 722225 (01)00884838099265(21); (6) Azurion 7820/15: 722068 (01)00884838085367(21) 722226 (01)00884838099272(21); (7) Azurion 7M12: 722078 (01)00884838085251(21) 722223 (01)00884838099241(21); (8) Azurion 7M20: 722079 (01)00884838085268(21) 722224 (01)00884838099258(21), Software: R2.2.O, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 |
Center Classification Date | 20230526 |
Recall Initiation Date | 20230321 |
Recalling Firm | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. |
Initial Notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
Similar To |