NRT X-RAY A/S Recall 92230
Description: Intelli-C, Left side suspended Tabletop, REF: 03400010
NRT X-RAY A/S Recall 92230 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1669-2023 |
Event ID | 92230 |
Event Description | Intelli-C, Left side suspended Tabletop, REF: 03400010 |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY. |
Quantity | 13 units |
Recall Reason | There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop. |
Device Classification | 20230607 |
Device Code Info | UDI-DI:05713464000121/Serial Numbers: 647009, 651002, 651006, 666009, 682003 |
Center Classification Date | 20230531 |
Recall Initiation Date | 20230427 |
Recalling Firm | NRT X-RAY A/S |