Terumo Cardiovascular Systems Corporation Recall 87834
Description: The Sarns TCM II (system) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. It also freezes water for an ice supply, monitors temperatures in the patient and extracorporeal circuit, and allows gradient rewarming relative to a venous blood temperature. The Sarns" TCM II also features a Cardioplegia System which will supply cooling water for cardioplegia. Device Name / Model Number: TCM II, with Cardioplegia, 110V/60Hz (P/N 164925) TCM II, without Cardioplegia, 110V/60Hz (P/N 164930) TCM II, with Cardioplegia, 220V/50Hz (P/N 164935) TCM II, without Cardioplegia, 220V/50Hz (P/N 164940) TCM (P/N 15747) Catalog Number: 4415, 164925, 164930, 164935, 15747
Terumo Cardiovascular Systems Corporation Recall 87834 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1791-2021 |
Event ID | 87834 |
Event Description | The Sarns TCM II (system) is a source of temperature-controlled water for blood heat exchangers used in an extracorporeal circuit and for blankets to externally heat or cool the patient. It also freezes water for an ice supply, monitors temperatures in the patient and extracorporeal circuit, and allows gradient rewarming relative to a venous blood temperature. The Sarns" TCM II also features a Cardioplegia System which will supply cooling water for cardioplegia. Device Name / Model Number: TCM II, with Cardioplegia, 110V/60Hz (P/N 164925) TCM II, without Cardioplegia, 110V/60Hz (P/N 164930) TCM II, with Cardioplegia, 220V/50Hz (P/N 164935) TCM II, without Cardioplegia, 220V/50Hz (P/N 164940) TCM (P/N 15747) Catalog Number: 4415, 164925, 164930, 164935, 15747 |
Product Type | Devices |
Distribution | Domestic: Foreign: Australia, Belgium, Canada, Chile, China, Colombia, Dominican Republic, England, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Philippines, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), Vietnam |
Quantity | 995 devices |
Recall Reason | Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices. |
Device Classification | 20210609 |
Device Code Info | All lot numbers distributed from 05/02/1985 thru 06/10/2015 |
Center Classification Date | 20210604 |
Recall Initiation Date | 20210416 |
Recalling Firm | Terumo Cardiovascular Systems Corporation |
Initial Notification | Letter |
Similar To |