Tandem Diabetes Care Inc Recall 90628
Description: t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology
Tandem Diabetes Care Inc Recall 90628 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1811-2022 |
Event ID | 90628 |
Event Description | t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology |
Product Type | Devices |
Distribution | US: NC, OK, ME, NY, TX, LA, NE, FL, AZ, IL, OH, CA, AL, MI, UT, VA, MA, SC, MS, GA, NV, HI, MN, WA, IN, AR, ID, CO, IA, MO, MT, PA, WI, NH, MD, WV, KS, NM, OR, KY, DE, ND, CT, NJ, TN, WY, AK, SD, DC, RI, IA, PR, VT, PT, VI, NC. OUS: Australia, New Zealand, South Africa |
Quantity | 31,100 |
Recall Reason | Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis. |
Device Classification | 20221005 |
Device Code Info | Pump/UDI-DI/Software Versions/Distributor Instructions: t:slim X2 insulin pump/00852162004781, 00852162004798, 00853052007240/ 4.3.5.2, 4.3.6/ For Distributors, All Parts, Contact Recalling Firm to Switch Out; t:slim X2 insulin pump with Dexcom G5/853052007257, 853052007271, 853052007295, 853052007318, 853052007967, 853052007981/ 5.0.1, 5.0.4, 5.3.1.1/ For Distributors, All Parts, Contact Recalling Firm to Switch Out; t:slim X2 insulin pump with Basal-IQ technology/853052007264, 853052007288, 853052007301, 853052007325, 850006613373, 850006613380, 850006613700, 850006613717, 850006613724, 850006613731/ 6.0.4, 6.3.1.1, 6.4/ For Distributors: 1) Parts: 1000886, 1000898, 1004219, 1005706, 1006373 or 1006375, Contact Recalling Firm to Switch Out, 2) Parts: 1010004, 1010005,1010006, No Action Required, Software is Up to Date; t:slim X2 insulin pump with Control-IQ technology/ 850006613229, 850006613205, 850006613212, 850006613762, 850006613779, 850006613786, 850006613793/ 7.3.1.1, 7.4/ For Distributors: 1) Parts: 1005011, 1005012, 1006402 or 1006404, Contact Recalling Firm to Switch Out, 2) Parts: 1010009,1010010, 1010011, No Action Required, Software is Up to Date |
Center Classification Date | 20220926 |
Recall Initiation Date | 20220524 |
Recalling Firm | Tandem Diabetes Care Inc |
Initial Notification | Letter |
Similar To |