Philips North America Llc Recall 94508
Description: SmartPath to dStream for 3.0T Model Number (REF): 782145
Philips North America Llc Recall 94508 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-1861-2024 |
Event ID | 94508 |
Event Description | SmartPath to dStream for 3.0T Model Number (REF): 782145 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom. |
Quantity | 3 units |
Recall Reason | The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire |
Device Classification | 20240529 |
Device Code Info | UDI-DI: N/A Serial Numbers: 17377 34114 38294 |
Center Classification Date | 20240517 |
Recall Initiation Date | 20240503 |
Recalling Firm | Philips North America Llc |
Initial Notification | Letter |