Leica Microsystems, Inc. Recall 90818

Description: Leica Surgical Operating Microscopes, ARveo 8, Part Number 10449063

Leica Microsystems, Inc. Recall 90818 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1848-2022
Event ID90818
Event DescriptionLeica Surgical Operating Microscopes, ARveo 8, Part Number 10449063
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity30 devices
Recall ReasonDuring internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
Device Classification20221005
Device Code InfoSerial Numbers: 30622001, 41221001, 100522001, 50322001, 60522001, 11221001, 301121001, 20622001, 191121001, 91221001, 61221001, 81221001, 281121001, 291121001, 181121001, 300522001, 71221001, 20322001, 30322001, 60322001, 61221002, 21221001, 241121001, 101221001, 251121001, 31221001, 261121001, 91221002, 101221002, 51221001,
Center Classification Date20220929
Recall Initiation Date20220826
Recalling FirmLeica Microsystems, Inc.