Philips North America Llc Recall 94508

Description: MR system 1.5T Marlin- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781483

Philips North America Llc Recall 94508 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1857-2024
Event ID94508
Event DescriptionMR system 1.5T Marlin- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781483
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Quantity10 units
Recall ReasonThe g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Device Classification20240529
Device Code InfoUDI-DI: N/A Serial Numbers: 79010 79008 79006 79009 79011 79003 79004 79002 79012 79005
Center Classification Date20240517
Recall Initiation Date20240503
Recalling FirmPhilips North America Llc