Adler MicroMed, Inc. Recall 87812

Description: Sapi Med sterile disposable colorectal anoscopes and proctoscopes. REF/Product Description: A.4083/THE BEAK - Surgical proctoscope. Disposable & sterile. Transparent. Self light; A.4519/MPR Short - Multi Purpose Rectoscope. 190mm long. Sterile & Disposable; A.4086/EPOFLIER - Surgical kit made of a 34,3mm wide anoscope and an opened retractor named GULL. Disposable & sterile; A.4084/THE BEAK Diagnostic- Disposable proctoscope. Transparent. Sterile & Self lighting; A.4522/ MPR - Multi Purpose Rectoscope. 220mm long. Sterile & Disposable

Adler MicroMed, Inc. Recall 87812 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1901-2021
Event ID87812
Event DescriptionSapi Med sterile disposable colorectal anoscopes and proctoscopes. REF/Product Description: A.4083/THE BEAK - Surgical proctoscope. Disposable & sterile. Transparent. Self light; A.4519/MPR Short - Multi Purpose Rectoscope. 190mm long. Sterile & Disposable; A.4086/EPOFLIER - Surgical kit made of a 34,3mm wide anoscope and an opened retractor named GULL. Disposable & sterile; A.4084/THE BEAK Diagnostic- Disposable proctoscope. Transparent. Sterile & Self lighting; A.4522/ MPR - Multi Purpose Rectoscope. 220mm long. Sterile & Disposable
Product TypeDevices
DistributionUS distribution: FL, IL, MA, and PA.
Quantity1,858 devices
Recall ReasonDue to issue with product sterility. Sterilization service provider had falsified records.
Device Classification20210630
Device Code InfoREF Numbers/Lot Numbers: REF A.4083 /Lot Numbers: 18D004l, 19C010l, 19l018l, 20l027l; REF A.4084/Lot Number: 19l015l; REF A.4519/Lot Numbers: 18l013l, 18l012l, 19D002l, 19F012l; REF A.4522/Lot Numbers: 18G003l, 19K023l; REF A.4086/Lot Number: 18F027l; Sterilization Batch Numbers: 18042311; 18092611; 18071211_1; 19040512; 18073112; 19092713; 19120612; 19063012; 20102212; 19063012;
Center Classification Date20210621
Recall Initiation Date20210423
Recalling FirmAdler MicroMed, Inc.
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit