Product Safety Recalls

Product Recall Tracker

Olympus Corporation of the Americas Recall 92325

Description: Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202LR.A

Olympus Corporation of the Americas Recall 92325 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-1987-2023
Event ID92325
Event DescriptionSingle Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202LR.A
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity340 pieces
Recall ReasonComplaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.
Device Classification20230628
Device Code InfoUDI-DI: 04953170385919 Lot Numbers: 24K, 27K, 29K, 2XK, 2YK
Center Classification Date20230621
Recall Initiation Date20230504
Recalling FirmOlympus Corporation of the Americas
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.