Permobil Recall 92250
Description: PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application
Permobil Recall 92250 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2004-2023 |
Event ID | 92250 |
Event Description | PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers: MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist Device via the SmartDrive MX2+ Application |
Product Type | Devices |
Distribution | Worldwide distribution. |
Quantity | 18,905 units |
Recall Reason | When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures. |
Device Classification | 20230705 |
Device Code Info | All units with software versions 1.0.00 to 1.0.04 |
Center Classification Date | 20230626 |
Recall Initiation Date | 20230510 |
Recalling Firm | Permobil |
Initial Notification | Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit |
Similar To |