Siemens Healthcare Diagnostics, Inc. Recall 88056
Description: Atellica IM 1300 Analyzer - multi-component system for in vitro diagnostic testing of clinical specimens SMN 11066001
Siemens Healthcare Diagnostics, Inc. Recall 88056 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2027-2021 |
Event ID | 88056 |
Event Description | Atellica IM 1300 Analyzer - multi-component system for in vitro diagnostic testing of clinical specimens SMN 11066001 |
Product Type | Devices |
Distribution | Nationwide Foreign: Country Name Algeria Argentina Australia Austria Bahrain Bangladesh Belgium Brazil Bulgaria Burkina Faso Canada Chile Colombia Croatia Cura¿ao Cyprus Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Greece Hungary India Indonesia Iraq Ireland Israel Italy Japan Kenya Kuwait Latvia Libya Lithuania Malaysia Mexico Morocco Netherlands New Zealand Norway Oman P.R. China Pakistan Peru Philippines Poland Portugal Qatar Republic Korea Romania Russian Fed. Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sweden Switzerland Taiwan Thailand Turkey U.A.E. Uganda Ukraine United Kingdom Uruguay Vatikancity Vietnam |
Quantity | US: 233 OUS: 233 |
Recall Reason | Valves used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject to high pressure or high wear and may impact the result accuracy of any assay. |
Device Classification | 20210714 |
Device Code Info | Effected Serial Range: IM00873 IM01452 and IRM00256xxxx IRM01169xxxx Note: xxxx in serial number is any number. UDI: 00630414007960 |
Center Classification Date | 20210707 |
Recall Initiation Date | 20210513 |
Recalling Firm | Siemens Healthcare Diagnostics, Inc. |
Initial Notification | Letter |