The Binding Site Group, Ltd. Recall 88064
Description: FREELITE Human Kappa Free Kit for use on the Roche cobas c Systems, Kappa Freie Leichtkette (Human) kit for Roche cobas c Systems, REF: LK016.CB, IVD CE, UDI: 05051700012096
The Binding Site Group, Ltd. Recall 88064 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2030-2021 |
Event ID | 88064 |
Event Description | FREELITE Human Kappa Free Kit for use on the Roche cobas c Systems, Kappa Freie Leichtkette (Human) kit for Roche cobas c Systems, REF: LK016.CB, IVD CE, UDI: 05051700012096 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide distribution in the states of CA, GA, FL, MA, MO, MI, WA, NY, PA, MN, AL, MD, OK, NE, IN, ID, OH, TX, HI, DC and the countries of Canada, Australia, Austria, Belgium, Bulgaria, Canada, Cyprus, Czech Republic, Denmark, Estonia, France, Germany, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovenia, Spain, Switzerland, United Kingdom, Argentina, Egypt, GUADELOUPE, HONG KONG, India, Iran, Israel, Korea, Lebanon, Malaysia, Morocco, Oman, Singapore, Taiwan, Tunisia, Vietnam. |
Recall Reason | The high and low level "Quality Controls" demonstrate a positive bias of approximately 10% compared to assigned values. |
Device Classification | 20210714 |
Device Code Info | Lot Numbers: 487719/ Exp: 31-Jul-2022, 480133/ Exp: 30-Sep-2022, Lot 457890/ Exp: 30-Sep-2022 |
Center Classification Date | 20210707 |
Recall Initiation Date | 20210520 |
Recalling Firm | The Binding Site Group, Ltd. |
Initial Notification | |
Similar To |