NeuroLogica Corporation Recall 92506
Description: OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck Model Number: NL5000
NeuroLogica Corporation Recall 92506 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2068-2023 |
Event ID | 92506 |
Event Description | OmniTom/OmniTom Elite- X-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck Model Number: NL5000 |
Product Type | Devices |
Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Canada, Grand Bahama, India, Italy, Korea, Portugal, Saudi, Scotland, Thailand. |
Quantity | 48 units |
Recall Reason | 1. Wheels loosening over time with use of the device could lead to the inability to effectively drive the device and could lead to delay in patient treatment. 2.Batteries used to power the NL5000 device contain contaminated boards, which may impact the ability to power on the device, this could lead to delay for patient treatment |
Device Classification | 20230712 |
Device Code Info | GTIN: 10815411020335 10815411020663 Serial Numbers: 10 14 23 25 26 36 38 40 41 42 44 46 48 49 50 51 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 15R |
Center Classification Date | 20230703 |
Recall Initiation Date | 20230606 |
Recalling Firm | NeuroLogica Corporation |