Fresenius Medical Care Holdings, Inc. Recall 87998

Description: Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are each intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician s prescription. Both concentrates are formulated for use in 45X proportioning systems which proportion a nominal ratio of 1 : 1.72 : 42.28 (acid : bicarbonate : water) to generate dialysate. The dialysate is intended to be pumped through a dialyzer, creating an osmotic gradient across the dialyzer membrane to exchange solutes with blood during hemodialysis.

Fresenius Medical Care Holdings, Inc. Recall 87998 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2086-2021
Event ID87998
Event DescriptionCitrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are each intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician s prescription. Both concentrates are formulated for use in 45X proportioning systems which proportion a nominal ratio of 1 : 1.72 : 42.28 (acid : bicarbonate : water) to generate dialysate. The dialysate is intended to be pumped through a dialyzer, creating an osmotic gradient across the dialyzer membrane to exchange solutes with blood during hemodialysis.
Product TypeDevices
DistributionUS Distribution to: TX only
Quantity(1) 1 case (2) 4 cases (3) 5 cases
Recall ReasonExposed to freezing temperatures due to refrigerated truck malfunctions
Device Classification20210728
Device Code InfoLot Number/UDI: (1)21BTAC004 UDI: (01)00840861101535 (2)21AGAC026 UDI: (01)00840861101696 (3)21BTAC008 UDI: (01)00840861101740
Center Classification Date20210716
Recall Initiation Date20210428
Recalling FirmFresenius Medical Care Holdings, Inc.