Fresenius Medical Care Holdings, Inc. Recall 87998
Description: Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are each intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician s prescription. Both concentrates are formulated for use in 45X proportioning systems which proportion a nominal ratio of 1 : 1.72 : 42.28 (acid : bicarbonate : water) to generate dialysate. The dialysate is intended to be pumped through a dialyzer, creating an osmotic gradient across the dialyzer membrane to exchange solutes with blood during hemodialysis.
Fresenius Medical Care Holdings, Inc. Recall 87998 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2086-2021 |
Event ID | 87998 |
Event Description | Citrasate Liquid Acid Concentrate and NaturaLyte Liquid Acid Concentrate are each intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician s prescription. Both concentrates are formulated for use in 45X proportioning systems which proportion a nominal ratio of 1 : 1.72 : 42.28 (acid : bicarbonate : water) to generate dialysate. The dialysate is intended to be pumped through a dialyzer, creating an osmotic gradient across the dialyzer membrane to exchange solutes with blood during hemodialysis. |
Product Type | Devices |
Distribution | US Distribution to: TX only |
Quantity | (1) 1 case (2) 4 cases (3) 5 cases |
Recall Reason | Exposed to freezing temperatures due to refrigerated truck malfunctions |
Device Classification | 20210728 |
Device Code Info | Lot Number/UDI: (1)21BTAC004 UDI: (01)00840861101535 (2)21AGAC026 UDI: (01)00840861101696 (3)21BTAC008 UDI: (01)00840861101740 |
Center Classification Date | 20210716 |
Recall Initiation Date | 20210428 |
Recalling Firm | Fresenius Medical Care Holdings, Inc. |
Initial Notification | Telephone |