Product Safety Recalls

Product Recall Tracker

Randox Laboratories Ltd. Recall 88130

Description: NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115

Randox Laboratories Ltd. Recall 88130 Information

StatusTerminated
Mandated?Voluntary: Firm initiated
Recall NumberZ-2099-2021
Event ID88130
Event DescriptionNEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115
Product TypeDevices
DistributionUS Nationwide Distribution in the states of AR, CA, CT, NY, OH, PA, TN, and TX
Quantity71 kits US
Recall ReasonStandard packed within NEFA kit has been assigned incorrectly. Calibrating with the incorrect value will cause a positive shift of up to +8% with both quality controls and patient samples
Device Classification20210728
Device Code InfoBatch Number: 544642 Exp. Date: 28th Nov 2022 GTIN: 05055273203066
Center Classification Date20210720
Recall Initiation Date20210510
Recalling FirmRandox Laboratories Ltd.
Initial Notification E-Mail
Similar To