Product Safety Recalls

Product Recall Tracker

GE OEC Medical Systems, Inc Recall 88202

Description: OEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures

GE OEC Medical Systems, Inc Recall 88202 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2160-2021
Event ID88202
Event DescriptionOEC 3D Imaging, Image-Intensified Fluoroscopic X-Ray System, Mobile - Product Usage: designed to provide fluoroscopic and digital spot images of adult and pediatric populations during diagnostic, interventional, and surgical procedures
Product TypeDevices
DistributionWorldwide distribution - US Nationwide distribution including in the states of CA, PA, WA, OH, LA, MS, AL, FL, MO, OK, AZ, NJ, CO, MI, NV, MN, KS, TN, NC, IL, MD, SC,GA, UT, WI, MT, AR, WV, ID,CT, IA, WY, VT, KY, DC, SD,ME, Virgin Islands and the countries of Colombia, Egypt, Peru, Chile, Spain, France, Germany, Hong Kong, Canada, Japan, Brazil, Turkey, United Arab Emirates, Singapore, Australia, Thailand, United Kingdom, Italy, Korea, Russia, CHINA, Romania, Nigeria, Czechia, Argentina, Taiwan, Bangladeh, Malaysia, Bosnia and Herzegovina, Poland, Belgium, Indonesia, Dominican Republic, Switzerland, Hungary, New Zealand, Vietnam, Mexico, India, Norway, Philippines, Ecuador, Lithuania, HONDURAS, Uruguay, Sweden, Croatia, Jamaica, Kuwait, Ethiopia, Saudi Arabia, Finland, Bahamas, Trinidad and Tobago, Pakistan, Slovakia, Costa Rica, Israel, Panama, Greece, South Africa and DENMARK.
Quantity3 units
Recall ReasonThere is a potential that the coin cell battery used to monitor X-Ray tube temperature may deplete early causing the system to become inoperable.
Device Classification20210804
Device Code InfoAll OEC 3D Systems
Center Classification Date20210729
Recall Initiation Date20210610
Recalling FirmGE OEC Medical Systems, Inc
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.