Product Safety Recalls

Product Recall Tracker

Boston Scientific Corporation Recall 92570

Description: Flexiva Pulse 242 TracTip Single-Use Laser Fibers -intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. 5-pack UPN: M006L8405961

Boston Scientific Corporation Recall 92570 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2230-2023
Event ID92570
Event DescriptionFlexiva Pulse 242 TracTip Single-Use Laser Fibers -intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy. 5-pack UPN: M006L8405961
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the country of Canada.
Quantity1245 units
Recall ReasonManufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched
Device Classification20230802
Device Code InfoGTIN: 08714729978664 Lot Numbers: 2030305543, 30305544, 30305545, 30305546, 30305547, 30305548, 30305549, 30389370, 30389371, 30389372, 30389373, 30389375, 30389376, 30389377, 30389378, 30429835, 30429837, 30479586, 30479587, 30479589, 30479590, 30521464, 30521466, 30521467, 30521468, 30554456, 30554457. Exp. Date: 7-Oct-2025 to 16-Nov-2025
Center Classification Date20230721
Recall Initiation Date20230525
Recalling FirmBoston Scientific Corporation