NEOSTEO Recall 88307

Description: Cannulated graduated drill bit 1.9 mm diameter, with AO QC system, REF: INS-114 for use in orthopedic surgery.

NEOSTEO Recall 88307 Information

StatusTerminated
Mandated?Voluntary: Firm initiated
Recall NumberZ-2236-2021
Event ID88307
Event DescriptionCannulated graduated drill bit 1.9 mm diameter, with AO QC system, REF: INS-114 for use in orthopedic surgery.
Product TypeDevices
DistributionUS Nationwide distribution in the state of New Jersey.
Quantity25
Recall ReasonTraces of corrosion on the laser marking was detected on cannulated graduated drill bit, which could lead to the emission of wear / corrosion particles in the operated area, potentially leading to the following risks: Inflammation, pain, allergy, adverse tissue reaction. Device description: The drill bit is used to create the hole required for the introduction of the screw, it is suitable for Neosteo compression screws for cortical and spongy bone. Drilling depth can be read on the instrument. The shaft is coupled with the ¿4.5mm surgical motor using a AO ¿4.5mm male drive connexion. The countersink bit is coupled to a ¿1.0mm k-wire to ensure guiding while drilling. The instrument is packaged in a plastic sheath sealed by welding.
Device Classification20210818
Device Code InfoLot: 243-20
Center Classification Date20210809
Recall Initiation Date20201103
Recalling FirmNEOSTEO
Initial Notification Letter