Simpleware Product Group, SYNOPSYS NORTHERN EUROPE Recall 88207
Description: Simpleware ScanIP software. For transfer of imaging information from a medical scanner to an output file, and for simulating/evaluating surgical treatment options.
Simpleware Product Group, SYNOPSYS NORTHERN EUROPE Recall 88207 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2253-2021 |
Event ID | 88207 |
Event Description | Simpleware ScanIP software. For transfer of imaging information from a medical scanner to an output file, and for simulating/evaluating surgical treatment options. |
Product Type | Devices |
Distribution | US: Nationwide Distribution in the states of CA, FL and countries of : France, United Kingdom, Germany, Australia |
Quantity | 4 softwares |
Recall Reason | A issue (bug) has been identified with the interface and image software which could result in anatomical orientation tags/labels being displayed incorrectly. This could result in operator misunderstanding and errors during resampling or re-alignment of image data |
Device Classification | 20210818 |
Device Code Info | Affected version/UDI Simpleware ScanIP 7.0: Created before UDI-DI system (released December 2nd, 2014), Simpleware ScanIP 2016.09: 00863520000307, Simpleware ScanIP M-2017.06: 00863520000314, Simpleware ScanIP N-2018.03: 00863520000321 |
Center Classification Date | 20210812 |
Recall Initiation Date | 20210604 |
Recalling Firm | Simpleware Product Group, SYNOPSYS NORTHERN EUROPE |
Initial Notification | Letter |