Zimmer, Inc. Recall 92681

Description: Zimmer Periarticular Locking Plate (ZPLP) Distal Lateral Femoral Plate 6 holes 159 mm Length-Indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, nonunions, and malunions. Item Number: 00235710106

Zimmer, Inc. Recall 92681 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2254-2023
Event ID92681
Event DescriptionZimmer Periarticular Locking Plate (ZPLP) Distal Lateral Femoral Plate 6 holes 159 mm Length-Indicated for temporary internal fixation and stabilization of osteotomies and fractures, including comminuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, nonunions, and malunions. Item Number: 00235710106
Product TypeDevices
DistributionAR AZ MA ME MO NJ NM NY OR WA Foreign: Brazil China Dominican Republic Korea Netherlamds Taiwan
Quantity75 units
Recall ReasonThread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.
Device Classification20230809
Device Code InfoLot Number/UDI: 64254511 /(01)0089024055940(10)64254511; 64296571/(01)0089024055940(10)64296571; 65379087/(01)0089024055940(10)65379087; 65379093/(01)0089024055940(10)65379093; 65379096/(01)0089024055940(10)65379096
Center Classification Date20230728
Recall Initiation Date20230627
Recalling FirmZimmer, Inc.
Initial Notification E-Mail
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.