American Contract Systems Inc Recall 94765
Description: ARTHROSCOPY SUP JOI PACK Procedure tray Catalog Number: BPAS17C
American Contract Systems Inc Recall 94765 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2281-2024 |
Event ID | 94765 |
Event Description | ARTHROSCOPY SUP JOI PACK Procedure tray Catalog Number: BPAS17C |
Product Type | Devices |
Distribution | US Nationwide distribution. |
Quantity | 272 units |
Recall Reason | Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012. |
Device Classification | 20240717 |
Device Code Info | UDI-DI: 00191072197825 LOT#"s: 665231 950241 |
Center Classification Date | 20240705 |
Recall Initiation Date | 20240515 |
Recalling Firm | American Contract Systems Inc |
Initial Notification | |
Similar To |