Product Safety Recalls

Product Recall Tracker

Baxter Healthcare Corporation Recall 94805

Description: Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.

Baxter Healthcare Corporation Recall 94805 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2323-2024
Event ID94805
Event DescriptionWelch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.
Product TypeDevices
DistributionDistribution to Illinois.
Quantity1 unit
Recall ReasonDuring manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.
Device Classification20240717
Device Code InfoUDI/DI: 00732094177640. Product Code: CP150A-1ENB. Serial Number: 100013892024
Center Classification Date20240710
Recall Initiation Date20240603
Recalling FirmBaxter Healthcare Corporation
Initial Notification Telephone
Similar To