Waldemar Link GmbH & Co. KG (Mfg Site) Recall 92758
Description: LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 17 - indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-117/00
Waldemar Link GmbH & Co. KG (Mfg Site) Recall 92758 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2348-2023 |
Event ID | 92758 |
Event Description | LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 17 - indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-117/00 |
Product Type | Devices |
Distribution | US Nationwide distribution. |
Quantity | 24 units |
Recall Reason | Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue |
Device Classification | 20230816 |
Device Code Info | UDI-DI: 04026575182114 Lot Numbers: B922124 B932051 C005064 C211011 |
Center Classification Date | 20230804 |
Recall Initiation Date | 20230703 |
Recalling Firm | Waldemar Link GmbH & Co. KG (Mfg Site) |
Initial Notification | |
Similar To |