Terumo Cardiovascular Systems Corporation Recall 88350
Description: Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
Terumo Cardiovascular Systems Corporation Recall 88350 Information
Status | Completed |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2350-2021 |
Event ID | 88350 |
Event Description | Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188 |
Product Type | Devices |
Distribution | International distribution in the country of Panama. |
Quantity | 1 unit OUS |
Recall Reason | intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs |
Device Classification | 20210901 |
Device Code Info | S/N: 01751 UDI: 00886799000588 |
Center Classification Date | 20210826 |
Recall Initiation Date | 20210608 |
Recalling Firm | Terumo Cardiovascular Systems Corporation |