Terumo Cardiovascular Systems Corporation Recall 88350

Description: Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188

Terumo Cardiovascular Systems Corporation Recall 88350 Information

StatusCompleted
Mandated?Voluntary: Firm initiated
Recall NumberZ-2350-2021
Event ID88350
Event DescriptionTerumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
Product TypeDevices
DistributionInternational distribution in the country of Panama.
Quantity1 unit OUS
Recall Reasonintermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs
Device Classification20210901
Device Code InfoS/N: 01751 UDI: 00886799000588
Center Classification Date20210826
Recall Initiation Date20210608
Recalling FirmTerumo Cardiovascular Systems Corporation