Product Safety Recalls

Product Recall Tracker

Ethicon Endo-Surgery Inc Recall 88434

Description: HARMONICHD 1000i Shears (36cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. Product Code: HARHD36

Ethicon Endo-Surgery Inc Recall 88434 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2356-2021
Event ID88434
Event DescriptionHARMONICHD 1000i Shears (36cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. Product Code: HARHD36
Product TypeDevices
DistributionWorldwide distribution - US Nationwide distribution and the countries of Austria, Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, United Arab Emirates.
Quantity68960 units
Recall ReasonAn internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns
Device Classification20210901
Device Code InfoLot Codes: U94Y9V U95C2A U94Z98 U95427 U9516E U95599 U9503G U95856 U94Y9W U95L0A U9518D U95526 U9507E U9548R U95127 U95525 U9530P U95T2Z U95366 U95R4F U9521Z U95F2T U9554U U95F8W U9564G U95U6A U95524 U95E6F U9512D U95K4D U94Z49 U95F07 U95A9X U95F6N U9569W U95E1U U9571P U95E72 U9587D U95E29 U95815 U95T90 U95814 U95Z67 U95754 U95Y7G U95D4T U9523L U95566 U9530X U9543R GTIN: 10705036015055
Center Classification Date20210826
Recall Initiation Date20210720
Recalling FirmEthicon Endo-Surgery Inc