Abbott Medical Recall 94798
Description: Abbott Proclaim XR 5 Implantable Pulse Generator, REF 3660, Sterile EO, Rx Only
Abbott Medical Recall 94798 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2358-2024 |
Event ID | 94798 |
Event Description | Abbott Proclaim XR 5 Implantable Pulse Generator, REF 3660, Sterile EO, Rx Only |
Product Type | Devices |
Distribution | Worldwide - US Nationwide distribution including in the states of ALABAMA, ALASKA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, GUAM, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA,MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, PUERTO RICO, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VERMONT, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN, WYOMING and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Ireland, Israel, Italy , Japan, Korea, Kuwait, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Poland, Portugal, Reunion Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom. |
Quantity | 83,511 units |
Recall Reason | The duration between the implantable pulse generator reaching the elective replacement indicator threshold and end of service may be shorter than indicated in product labeling. |
Device Classification | 20240724 |
Device Code Info | UDI: 05415067031419/ All Serial Numbers |
Center Classification Date | 20240716 |
Recall Initiation Date | 20240516 |
Recalling Firm | Abbott Medical |
Initial Notification | Letter |