Covidien LP Recall 92660
Description: MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions REF: 8888345520HP
Covidien LP Recall 92660 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2383-2023 |
Event ID | 92660 |
Event Description | MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions REF: 8888345520HP |
Product Type | Devices |
Distribution | US Nationwide - Worldwide Distribution Foreign: Argentina Austria Bahrain Bangladesh Belgium Brazil Brunei Darussalam Canada Cayman Islands China Costa Rica Czech Republic Denmark Dominican Republic Finland France French Polynesia Germany Greece Hungary India Iran, Islamic Republic Of Ireland Italy Kuwait Malaysia Maldives Mauritius Mexico Nepal Netherlands New Caledonia Nicaragua Norway Philippines Poland Portugal Puerto Rico Saudi Arabia Singapore Slovakia South Africa Spain Sri Lanka Sweden Switzerland Thailand Turkey United Arab Emirates United Kingdom |
Quantity | 295 units |
Recall Reason | Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus |
Device Classification | 20230823 |
Device Code Info | GTIN: 20884521128115 LOT Numbers: 2018800011 2203900129 |
Center Classification Date | 20230811 |
Recall Initiation Date | 20230628 |
Recalling Firm | Covidien LP |
Initial Notification | Letter |
Similar To |