Product Safety Recalls

Product Recall Tracker

B. Braun Medical, Inc. Recall 94823

Description: Infusomat 60DROP METRISET PUMP SET, 3 SAFELINE-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 362034

B. Braun Medical, Inc. Recall 94823 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2380-2024
Event ID94823
Event DescriptionInfusomat 60DROP METRISET PUMP SET, 3 SAFELINE-Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/Catalog Number: 362034
Product TypeDevices
DistributionWorldwide distribution - US Nationwide and the country of Canada.
Quantity280 units
Recall ReasonPotential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
Device Classification20240724
Device Code InfoUDI-DI 04046964182570; All Lots Distributed After 10AUG2023.
Center Classification Date20240717
Recall Initiation Date20240617
Recalling FirmB. Braun Medical, Inc.
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.