Carefusion 2200 Inc Recall 92561

Description: Snowden-Pencer MIS DIAMOND-TOUCH CLAMP BOWEL 50 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7045; endoscopic surgical device

Carefusion 2200 Inc Recall 92561 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2421-2023
Event ID92561
Event DescriptionSnowden-Pencer MIS DIAMOND-TOUCH CLAMP BOWEL 50 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7045; endoscopic surgical device
Product TypeDevices
DistributionUS and Canada
Quantity2 units
Recall ReasonSpecific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.
Device Classification20230823
Device Code InfoUDI/DI 10885403160561, Lot Number B23
Center Classification Date20230817
Recall Initiation Date20230713
Recalling FirmCarefusion 2200 Inc