LINK BIO CORP Recall 94926

Description: LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00

LINK BIO CORP Recall 94926 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2445-2024
Event ID94926
Event DescriptionLINK SymphoKnee Patella Sizing Template, Item Number 881-509/00
Product TypeDevices
DistributionUS Nationwide distribution in the states of FL, GA, KS, OH, WI.
Quantity8 units
Recall ReasonThe LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process.
Device Classification20240731
Device Code InfoItem Number: 881-509/00; UDI-DI: 04026575310203; Lot number: C225066.
Center Classification Date20240725
Recall Initiation Date20240627
Recalling FirmLINK BIO CORP
Initial Notification E-Mail