LINK BIO CORP Recall 94926
Description: LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00
LINK BIO CORP Recall 94926 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2445-2024 |
Event ID | 94926 |
Event Description | LINK SymphoKnee Patella Sizing Template, Item Number 881-509/00 |
Product Type | Devices |
Distribution | US Nationwide distribution in the states of FL, GA, KS, OH, WI. |
Quantity | 8 units |
Recall Reason | The LinkSymphoKnee Patella Sizing Template was incorrectly labeled. The markings on two arms of the instrument were mixed up because the triangular Patella Sizing Template was inserted upside down into the laser marking process. |
Device Classification | 20240731 |
Device Code Info | Item Number: 881-509/00; UDI-DI: 04026575310203; Lot number: C225066. |
Center Classification Date | 20240725 |
Recall Initiation Date | 20240627 |
Recalling Firm | LINK BIO CORP |