Product Safety Recalls

Product Recall Tracker

Siemens Medical Solutions USA, Inc Recall 88588

Description: MR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815

Siemens Medical Solutions USA, Inc Recall 88588 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2447-2021
Event ID88588
Event DescriptionMR MAGNETOM Systems with RT Image Suite with syngo MR VA20A, VA30A and VA31A with license SW-Key CT_RT_IMA_SUITE_SYNTHETIC_CT Material Numbers: 11060815
Product TypeDevices
DistributionUS Nationwide distribution.
Quantity2 systems
Recall ReasonPotential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).
Device Classification20210922
Device Code InfoSerial Numbers: 176302 175681
Center Classification Date20210910
Recall Initiation Date20210816
Recalling FirmSiemens Medical Solutions USA, Inc
Initial Notification Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Similar To

© 2021 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.