Hamilton Medical, Inc. Recall 92873
Description: Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator
Hamilton Medical, Inc. Recall 92873 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2529-2023 |
Event ID | 92873 |
Event Description | Hamilton C1/T1/MR1, REF: MSP161502/12, Control Board Spare part for ventilator |
Product Type | Devices |
Distribution | Distribution US nationwide, including Puerto Rico. OUS: Not provided |
Quantity | 6 units |
Recall Reason | Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation. |
Device Classification | 20230927 |
Device Code Info | UDI: 07630002805247 / Affected Serial Numbers: 4001901, 4002139, 4003445, 4003567, 4003576, 4005595 |
Center Classification Date | 20230920 |
Recall Initiation Date | 20230726 |
Recalling Firm | Hamilton Medical, Inc. |
Initial Notification | Letter |