Landauer Recall 92647
Description: nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR readers
Landauer Recall 92647 Information
Status | Ongoing |
Mandated? | Voluntary: Firm initiated |
Recall Number | Z-2553-2023 |
Event ID | 92647 |
Event Description | nanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR readers |
Product Type | Devices |
Distribution | Worldwide |
Quantity | 16760 units |
Recall Reason | LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure. |
Device Classification | 20230920 |
Device Code Info | UDI/DI 0860003399903, all batch numbers |
Center Classification Date | 20230912 |
Recall Initiation Date | 20230712 |
Recalling Firm | Landauer |
Initial Notification | Letter |
Similar To |