Product Safety Recalls

Product Recall Tracker

Landauer Recall 92647

Description: nanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO; d) 03057-SMO; radiation monitoring dosimeter used with the microSTAR readers

Landauer Recall 92647 Information

StatusOngoing
Mandated?Voluntary: Firm initiated
Recall NumberZ-2544-2023
Event ID92647
Event DescriptionnanoDot D2DNS, Model Numbers: a) 03057-1MO; b) 03057-3MO; c) 03057-OTO; d) 03057-SMO; radiation monitoring dosimeter used with the microSTAR readers
Product TypeDevices
DistributionWorldwide
Quantity18638 units
Recall ReasonLANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
Device Classification20230920
Device Code InfoUDI/DI 0860003399903, all batch numbers
Center Classification Date20230912
Recall Initiation Date20230712
Recalling FirmLandauer
Initial Notification Letter
Similar To

© 2022 Recall.Report / Contact / Privacy Policy

This site's data is aggregated programmatically and provided "as is" without any representations or warranties, express or implied.
Recall.report is not affiliated with the Consumer Product Safety Commision, FDA, or US government in any way.